KES diode laser hair removal handpiece repair Description

This VS-MI-100-6-FL-B-808 is an 808nm diode laser stack, which has 6pcs 100w bars with PEEK shell and backing board. Electrodes are on the back of the stack

It is used for KES diode laser hair removal handpiece repairing and replacement.

Oriental-laser focused on diode laser bar assembling for more than 16 years and we have seen various alma soprano diode laser stacks.Also we can use the same Jenoptik bar like Alma used to recondition the power the laser soprano power.


KES Laser handpiece Repairing procedure

>Choice A: Send the old broken handpiece to our factory. We will do the process of checking , repairing, cleaning and etc. Then send the fresh new handpiece back to you.

>Choice B:  if you have basic assembly capacity, you can send the diode laser stack only to our factory. We remove the old bar and mount the new bar to recover the laser power of the stack. 

>Choice C: Measure your stack and confirm the water pipe hole and screw hole position. Then press the button below to buy a ready-made stack which is match your old handpiece. 


KES Laser handpiece Advantages

>Save Money. Replace the broken diode stack only. Low cost and convenient

>Long lifetime.  we use diode laser bars from Germany. The newest Ausn mounting technology will be applied when repairing. After repair. The handpiece will be as a new one. lifetime is much longer than before

>All technical support available. We major in diode laser mounting, software, spare parts over 16 years. All the technologies about laser hair removal machine are open to you.           


KES diode laser stack Specification   

    

Model No. 

VS-MI-100-6-FL-B-808

Application

laser hair removal Handpiece Repair And Replacement  exp: KES

Operation Mode

CW

Power per submounts

100W

Number of submounts in one stack

6bars

Central wavelength at 25℃

808 ± 5nm

Typical Operation current

85A

Maximum Operating current

100A

Typical Threshold current

15A

Maximum Operating Voltage

<12V

Fast Axis Divergence (FWHM)

39º

Slow Axis Divergence (FWHM)

Operation Condition

non-condensing atmosphere, cleaning room

Expected lifetime

10000 hours

Typical pitch per bar

1.8mm

Cooling requirement


Flow rate

1.8L/min +-10%

Water Temperature

25+-3°C

Water Quality

Deionized 2-10 µs/cm, mixed bed ion exchanger,particle filter<25µm

cooling system

Do not use any material that combination with copper would form galvanic elements